Abstract

The quantification of albumin is important in clinical medicine because the concentration of albumin in biological fluids is closely related to human health. In this study, we developed a highly selective and robust assay to determine human serum albumin (HSA) in human plasma by combining chymotrypsin/trypsin digestion coupled with targeted LC-MS/MS technique. Human plasma samples were denatured, reduced, alkylated, and digested with both chymotrypsin and trypsin to generate surrogate peptides. A unique chymotryptic peptide (NAETF) arising from human serum albumin was finally selected for targeted LC-MS/MS detection and quantification. Numerous parameters related to the targeted LC-MS/MS assay were evaluated, including lower limit of quantitation (LLOQ), linearity range, enzyme digestion efficiency, accuracy and precision. The LC-MS/MS assay was linear in the concentration range 0.05–1 mg/mL with intra-day and inter-day precision <10.2% and accuracy ranging from −3.94% to 4.89%. The assay was successfully applied to determine HSA in 148 human plasma samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.