Abstract

Analytical method for an accurate determination of folic acid in human serum has been established using isotope dilution ultra‐high‐pressure liquid chromatography/mass spectrometry (ID‐UPLC/MS). Sample extract was prepared from 0.5 mL of human serum spiked with 13C5‐folic acid solution and then LC–MS/MS experiment was performed on multiple reaction monitoring mode for 20 min. The method was tested for repeatability using homogeneous human serum and it showed good repeatability with relative standard deviation of 1.8%. Further evaluation was carried out through NIST SRM 1955 samples (level I, II, and III) and the measurement results were in good agreement with the certified values with good reproducibility over two periods. It was also revealed that the obtained relative expanded uncertainty was 12% which is much smaller than that of previously reported methods, indicating that this method is very reliable and accurate for determining the content of folic acid in human serum.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.