Abstract

BackgroundOur aim was to evaluate the reproducibility of epicardial adipose tissue (EAT) volume, measured on scans performed using an open-bore magnetic resonance scanner.MethodsConsecutive patients referred for bariatric surgery, aged between 18 and 65 years who agreed to undergo cardiac imaging (MRI), were prospectively enrolled. All those with cardiac pathology or contraindications to MRI were excluded. MRI was performed on a 1.0-T open-bore scanner, and EAT was segmented on all scans at both systolic and diastolic phase by two independent readers (R1 with four years of experience and R2 with one year). Data were reported as median and interquartile range; agreement and differences were appraised with Bland-Altman analyses and Wilcoxon tests, respectively.ResultsFourteen patients, 11 females (79%) aged 44 (41–50) years, underwent cardiac MRI. For the first and second readings, respectively, EAT volume was 86 (78–95) cm3 and 85 (79–91) cm3 at systole and 82 (74–95) cm3 and 81 (75–94) cm3 at diastole for R1, and 89 (79–99) cm3 and 93 (84–98) cm3 at systole and 92 (85–103) cm3 and 93 (82–94) cm3 at diastole for R2. R1 had the best reproducibility at diastole (bias 0.3 cm3, standard deviation of the differences (SD) 3.3 cm3). R2 had the worst reproducibility at diastole (bias 3.9 cm3, SD 12.1 cm3). The only significant difference between systole and diastole was at the first reading by R1 (p = 0.016). The greatest bias was that of inter-reader reproducibility at diastole (-9.4 cm3).ConclusionsReproducibility was within clinically acceptable limits in most instances.