Abstract

We developed a handheld fluorescence detection system for capillary-based enzyme-linked immunosorbent assay (ELISA). The detection system implements both a long-pass filter and perpendicular optical arrangement, i.e., a power LED and a palm-sized spectrometer, to minimize background signals from the excitation light and optical scattering. The lower detection limit for resorufin was 0.13 μM. The detection system was applied to the quantification of C-reactive protein (CRP) in human serum with a capillary-based ELISA. The lower detection limit for CRP was 31 ng/ml, and the observed CRP levels in human serum were comparable to those obtained with a conventional ELISA system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.