Abstract

A new technique for measuring urethral opening pressure above the distal sphincter is described, based on the fluid bridge test. This test eliminates many of the inaccuracies produced when using standard pressure flow studies. Nineteen males were studied, 15 of whom had cystometric evidence of bladder outlet obstruction, the other 4 being unobstructed. The mean urethral opening pressure in the obstructed group was 69 cm H2O (range 44-124) and in the unobstructed group 19 cm H2O (range 16-22). The test is simple to perform and is not affected by muscular activity in the pelvic floor or the time lag between detrusor pressure and flow rate measurement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.