Abstract

In the present work, a permanent magnet-based helicon plasma source (HELEN) is characterized as a negative ion source. A noninvasive diagnostic technique based on cavity ring down spectroscopy (CRDS) is developed to measure the line-integrated negative hydrogen ion (H-) density in HELEN. This paper discusses the experimental results in which negative hydrogen ion density is measured for different magnetic field, pressure, and RF (13.56 MHz) power configurations. It is observed that in high power range (800-850 W), the source is operating in helicon wave heating mode with a background plasma density of ∼1018 m-3 and electron temperature of ∼2 eV. Negative hydrogen ion density is measured by CRDS method and also by optical emission spectroscopic technique. The measured negative ion density is in the range of ∼1016 m-3 under volume mode operational condition even without any use of standard magnetic filter or cesium seeding in the downstream region. The influence of pressure variation (pressure range from 4 × 10-3 mbar to 4 × 10-2 mbar) on negative ion density production is not significant, except for a particular axial magnetic field configuration (55 G) and at 8 × 10-3 mbar pressure, where ∼34% hike is observed.