Abstract

An online dynamic pH junction preconcentration method was developed for quantification of 5-aminolevulinic acid (ALA) by CE with the separation time less than 6 min. The optimal dynamic pH junction of ALA was carried out between pH 9.3 borate buffer (BGE, 40 mM) and pH 2.5 phosphate buffer (sample matrix, 40 mM) when 4.1 cm of sample plug was hydrodynamically injected into an uncoated fused-silica capillary (48.5 cm in length, id of 50 μm). If a 24 kV separation voltage was applied, the calibration curve of ALA peak area (200 nm) showed good linearity (R(2) = 0.9991) ranging from 0.01 to 6.5 mg/mL. The reproducibility of this system was excellent with RSDs (n = 10) of 2.5% for peak area response and 0.6% for migration time at ALA concentration of 0.5 mg/mL. The LOD was evaluated as 1.0 μg/mL (S/N > 3). Compared to conventional CE procedure, the sensitivity was successfully improved over 50-fold. The analytical results of pharmaceutical formulations show a good agreement with those by HPLC (r = 0.94).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.