Abstract

We built a heating furnace using stainless-steel instead of aluminum in gas chromatography combined with an oxidation/reduction system; it increased the oxidation temperature to 650°C. At 600°C, it completely oxidized five organochlorine compounds. This system was applied to a standard solution of 23 volatile organic compounds. The analytical results of 20 hydrocarbon and organochlorine compounds showed good agreement with the expanded uncertainty (k = 2) of the reference values. Three organobromine compounds obtained values higher than the reference; this was investigated further.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.