Abstract

In order to evaluate density, radiographic contrast and dose of radiation exposure, the author analyzed 80 radiographs containing 640 optical density data of the images of a penetrometer, exposed to the radiation beam with combinations between D and E periapical films, aluminum and copper/aluminum filters, and circular or rectangular collimators. The data obtained were analyzed by ANOVA and allowed the following conclusions: 1) aluminum filtration resulted in improved image contrast; 2) the use of group D film and an aluminum filter produced improved image contrast quality; 3) the rectangular collimator contributed to the production of improved contrast and to the reduction of radiation exposure, but did not affect density; 4) the combination of copper/aluminum filter, E group film and rectangular collimation significantly reduced radiation exposure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.