Abstract

Quality evaluation of the k0-standardized neutron activation analysis (k0-NAA) at Dalat research reactor (DRR) has been implemented by analyzing the SMELS, a synthetic multi-element standard material as well as the certified reference materials for purposes of validation and verification of the method. The results revealed that the k0-NAA established at DRR has met the requirements of multi-element analysis for the intended applications with about 42 elements: Al, As, Au, Ba, Br, Ca, Ce, Cl, Co, Cr, Cs, Cu, Dy, Eu, Fe, Hf, I, In, K, La, Mg, Mn, Mo, Na, Nd, Pr, Rb, Sb, Sc, Se, Sm, Sr, Ta, Tb, Th, Ti, U, V, W, Yb, Zn and Zr in matrices of environment, geology and biology among others. Evaluation criteria for the laboratory quality used in this work are the statistical analysis parameters namely: u-scores, laboratory result/certified value ratios and relative biases as well as the control charts those are shown in the paper.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.