Abstract

Quality control helps radio-diagnostic facilities achieve appropriate radiological information with little dose and cost. In this study, quality control tests included Kilovoltage kVp accuracy and reproducibility, KVp and timer reproducibility, taking voltage verses radiation doses checking and reproducibility, milliampere(mA) and timer linearity, milliampere second mAs reproducibility and output doses verses mAs, light-radiation collimator integrity, Perpendicularity, and HVL These results were compared to established methods. A certified physicist should do this test. Equipment functionality was determined via the status test. This was done shortly following the acceptance test. The test is repeated after repair that affects functional status, like the acceptance test. A certified physicist performs the status test, which involves absolute measurements. All processes at AECH DINAR D.I. Khan were within acceptable limits.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.