Abstract
Custom blocks are the most commonly used field shaping device in radiotherapy. They offer many advantages in daily clinical use. However, as with all accessories used for patient treatments, care must be taken with their use. As such, a quality control program to check block cutting accuracy is recommended as a routine part of their clinical use. We have developed a three step method for checking custom blocks before they are used for patient treatment. These steps include: a static light check, a parallel opposed film check, and a block check involving the patient. By completing these three steps, we feel that we have improved the overall accuracy of our custom block making system which has resulted in more accurate treatments for our patients.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.