Abstract

The Digital Mammography Imaging Screening Trial (DMIST), conducted under the auspices of the American College of Radiology Imaging Network (ACRIN), is a clinical trial designed to compare the accuracy of digital versus screen-film mammography in a screening population [E. Pisano et al., ACRIN 6652-Digital vs. Screen-Film Mammography, ACRIN (2001)]. Part I of this work described the Quality Control program developed to ensure consistency and optimal operation of the digital equipment. For many of the tests, there were no failures during the 24 months imaging was performed in DMIST. When systems failed, they generally did so suddenly rather than through gradual deterioration of performance. In this part, the utility and effectiveness of those tests are considered. This suggests that after verification of proper operation, routine extensive testing would be of minimal value. A recommended set of tests is presented including additional and improved tests, which we believe meet the intent and spirit of the Mammography Quality Standards Act regulations to ensure that full-field digital mammography systems are functioning correctly, and consistently producing mammograms of excellent image quality.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.