Abstract

Analytical Quality by Design principles using the design of experiments were applied for the development of a capillary electrophoresis method for the determination of enantiomeric purity and chemically related impurities of tamsulosin. From initial scouting experiments, a dual cyclodextrin (CD) system composed of sulfated β-CD and carboxymethyl-α-CD was selected as the chiral selector. A fractional factorial resolution V+ design was used for the identification of the critical process parameters, while a face-centered central composite design and Monte Carlo simulations were employed for final optimization and defining the design space of the method. The experimental conditions of the working point were: 30mM sodium phosphate buffer, pH 3.0, containing 40mg/mL sulfated β-CD and 7mg/mL carboxymethyl-α-CD, capillary temperature 18°C, applied voltage -23kV. Following the assessment of robustness by applying a Plackett-Burman design, the method was validated according to the International Council for Harmonization of Technical Requirements for Pharmaceuticals for Human Use guideline Q2(R1). The method allowed the quantification of the chiral impurity and three other related impurities at the 0.1 % level with acceptable accuracy and precision.