Abstract
Gating response times of monitoring system/LINAC combinations for gated radiotherapy treatments are notoriously difficult to measure. This difficulty may reflect why the incorporation of gating response times is not thoroughly considered when creating treatment margins for gated radiotherapy treatments, however ignoring the effect of gating response time could lead to significant treatment inaccuracies. This study shows a methodology which measured gating response times for AlignRT/Varian Truebeam combination which appears to be applicable to any surface guided monitoring/Linac combination as well as those combinations which incorporate extended lag times (>200 ms). Beam on lag time appears to be measurably greater than beam off lag time. Although these gating response times are slower than other gating systems on the market, the advantages surface guided radiotherapy (SGRT) could potentially provide for treatment accuracy is discussed as well as demonstrating the lack of guidance regarding SGRT with respect to gated treatments.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.