Abstract

For the acceptance test and annual quality assurance of the Leksell Gamma Unit, measurement of the beam accuracy, defined as a distance between mechanical and radiological isocenters, poses a challenge to medical physicists. The specification for the beam accuracy is within 0.5 mm for the 4‐mm collimator helmet. In this report, we introduce a simple technique to analyze the beam accuracy by using a conventional film densitometer plus mathematical modeling. A small piece of film was placed inside the film cassette containing a sharp needle. The needle is located such that its tip is exactly positioned at the mechanical isocenter. Before exposure, the film was pierced by the needle. Density profile was measured by using a densitometer with a spatial resolution of 0.8 mm. The profile was then fitted to a model of the two Gaussian functions. One is for the radiation field profile, the other for a dip caused by the narrow hole. The difference between the centers of the two Gaussian functions defines the deviation of the beam accuracy from the mechanical center of the unit. The deviations for x, y, and z directions from one of our annual measurements are 0.032, 0.054, and 0.195 mm, respectively. The combined deviation is 0.20 mm, which is well within the specification and in excellent agreement with the results from the manufacture's laser measurement. This technique provides a simple, accurate and practical tool for measurement of the beam accuracy in the acceptance test and annual quality assurance of the Leksell Gamma Unit.PACS number(s): 87.66.–a, 87.53.–j

Highlights

  • In stereotactic radiosurgery, such as Leksell Gamma Knife radiosurgery, a single high dose of radiation is delivered to a radiological lesion by using multiple high accurately focused beams

  • For the Leksell Gamma Unit, the beam accuracy is defined as a distance or deviation between mechanical and radiological isocenters

  • We introduce a simple technique to analyze the beam accuracy of the Leksell Gamma UnitModel Udeveloped at USC University Hospital by using a conventional film densitometer plus mathematical modeling

Read more

Summary

Quality Assurance of Beam Accuracy for Leksell Gamma Unit

Cheng Yua) and Zbigniew Petrovich Department of Radiation Oncology, University of Southern California, School of Medicine, 1441 Eastlake Avenue, Los Angeles, California 90033 Gary Luxtonb) Department of Radiation Oncology, Stanford University School of Medicine, Stanford, California 94305

INTRODUCTION
METHODS
RESULTS AND DISCUSSIONS
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.