Abstract

In vitro testing of blood contacting materials before clinical application is generally advisable. Four heparin coatings from different manufacturers were tested for adsorbed proteins and soluble activation markers. The surface with the highest antithrombin, thrombin, high-molecular-weight-kininogen (HMWK) and the lowest fibrinogen binding capacity (Carmeda, Medtronic) showed significantly lower levels of granulocytes and platelet activation (beta-TG, PMN-elastase release). No statistically significant differences in soluble markers of the coagulation system could be detected (F1 + 2, TAT). Interestingly, complement activation (TCC) was significantly reduced within the group of the lowest adsorption of the complement factor C3. Our data demonstrate that there is a relation between the binding affinity of proteins (C1-inhibitor, C3-complement) and the consecutive changes in complement activation (TCC). Therefore, measuring adsorbed proteins on artificial surfaces is a suitable, sensitive and very reproducible method for assessing the thrombogenicity of biomaterials.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.