Abstract

Comparison of dosage and spatial resolution of digital X-Ray devices with 3D-capability in head and neck imaging. Three on-site X-Ray devices, a general purpose multi-slice CT (CT), a dedicated cone-beam CT (CBCT) and the CT-mode of a device for digital angiography (DSA) of the same generation were compared using paranasal sinus (PNS) and temporal bone imaging protocols. The radiation exposure was measured with a puncture measuring chamber on a CTDI head phantom as well as with chip-strate-dosimeters on an Alderson head phantom in the regions of the eyes and thyroid gland. By using the Alderson head phantom, the specific dosage of the X-Ray device with regard to different protocols was read out. For the assessment of the high-contrast resolution of the devices, images of a self-made phantom were qualitatively analysed by six observers. The three devices showed marked variations in the dosage and spatial resolution depending on the protocol and/or modus. In both parameters, CBCT was superior to CT and DSA using standard protocols, with the difference being less obvious for the investigation with PNS. For high-contrast investigations CBCT CT is a remarkable option in head and neck radiology.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.