Abstract

A recently introduced rapid-scanning quadrupole mass spectrometer (qMS) with an electron-capture negative ion (ECNI) option, the Perkin-Elmer Clarus 500, was tested as a detector for comprehensive two-dimensional gas chromatography (GC × GC). The parameters influencing the data acquisition rate in the scan mode, such as scan time and inter-scan delay, and in the selected ion monitoring mode, such as dwell time and inter-channel delay, were evaluated. In the scan mode, good-quality mass spectra covering a range of 300 Da can be obtained at an acquisition rate of 23 Hz; in selected ion monitoring, an acquisition rate of 90 Hz can be achieved when monitoring a single ion. Compared with electron ionisation, the use of electron-capture negative ionisation causes no extra peak broadening. As applications, mixtures of polychlorinated n-alkanes (PCAs), polybrominated diphenyl ethers (PBDEs) and polychlorinated dibenzo- p-dioxins (PCDDs) and dibenzofurans (PCDFs) were analysed. The separation of PCAs based on their number of chlorine substituents was confirmed for the first time by using GC × GC–ECNI qMS in the scan mode and a significantly improved limit of detection was achieved for BDEs (10–150 fg injected) and CDD/Fs (10–700 fg injected) in the selected ion monitoring mode.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.