Abstract
The aim of the study is to assess the performance of an automated method for Q-onset and T-end delineation, as well as QT measurement with the use of a ‘gold standard’ of a manually created reference database. The data to be used comprise 548 recordings of the PTB Diagnostic ECG Database. The ECG signals are preprocessed, suppressing power-line interference, electromyographic noise and baseline drift according to our previously published investigations, guaranteeing accurate Q-onset and T-end locality preservation. Our method for automatic detection of Q-onset and T-end is based on the minimum value of the angle between two segments having a common mid point and equal lengths of 10 ms. The minimum of the angle is searched in defined time intervals delimited separately for the Q-onset and T-end. All measurements are performed on lead II only. Mean ± standard deviations are obtained for 95% of the recordings: −0.08 ± 2.71 for Q-onset, 5.10 ± 9.22 for T-end and 4.40 ± 9.93 for QT interval, as well as for 100% of the recordings: 0.46 ± 4.84 for Q-onset, 1.28 ± 16.75 for T-end and 0.83 ± 16.67 for QT interval.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.