Abstract

Abstract A micro-pump for medical applications is being developed using organometallic derived PZT thin films and silicon micromachining technology. Critical issues include the deposition of ‘thick’ PZT films (1 to 2 μm thickness), film processing compatibility with available Si/SixNy/Ta/Pt structures and the establishing of piezoelectric poling parameters. A processing technique for the preparation of thick PZT films is presented. Piezoelectric properties are characterized by both direct method using alternating compressional force and by converse method using laser interferometry. In poled films of approximately 1. 0 μm thickness, d33 values of about 130 pC/N were measured by both direct and converse methods. Maximum d33 constants of ≈ 200 pC/N were measured with 8 V bias voltage. Mechanical resonance was excited in a 1 cm by 1 cm silicon substrate by a 0.8 μm thick PZT film poled at 15 V.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.