Abstract

Currently, one of the main metrological noncontact temperature measurement challenges is the emissivity uncertainty. This paper describes a pyrometer with emissivity effect diminishing through the use of a measuring scheme with tracking balancing in which the radiation receiver is a null-indicator. In this paper the results of the prototype pyrometer absolute error study in surfaces temperature measurement of aluminum and nickel samples are presented. There is absolute error calculated values comparison considering the emissivity table values with errors on the results of experimental measurements by the proposed method. The practical implementation of the proposed technical solution has allowed two times to reduce the error due to the emissivity uncertainty.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.