Abstract

An antibody-based, label-free impedance biosensor for the sepsis biomarker procalcitonin (PCT) has been produced using highly oriented pyrolytic graphite (HOPG) as the sensor electrode material and with attachment of the PCT capture antibody via coupling to 1-pyrenebutanoic acid succinimidyl ester (PYSE) pre-immobilised on the electrode by π-π stacking. In an aqueous phosphate-buffered saline (PBS) at pH = 7.4, the charge-transfer resistance for the redox probe Ru(NH3)6 3+/2+ exhibited a linear dependence on the PCT concentration in the range 10 to 100 ng mL−1 and provided a detection limit for PCT of 3.70 ng mL−1. Also, over the same PCT concentration range, the electrode capacitance showed a steady decrease, offering the prospect of a capacitive sensing mode for the label-free device, possibly removing the need to add a redox species to the sample under test. These results demonstrate the feasibility of using pyrene-modified HOPG as a simple platform for label-free impedance biosensors, here for the important sepsis biomarker PCT.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.