Abstract

There is dispute about the validity of the nitrogen rebreathing technique for determination of Pv̄ O 2 . This theoretical analysis examines the circumstances under which the technique is likely to be applicable. When the plateau method is used the probable error in Pv̄ O 2 is ±2.5 mm Hg at rest, and of the order of ±1 mm Hg during exercise. Provided the rebreathing bag size is reasonably chosen, Denison's extrapolation technique gives results at least as accurate as those obtained by the plateau method. At rest, however, extrapolation should be to 30 rather than to 20 sec.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.