Abstract

A purge-and-trap method for biological monitoring of styrene in urine was developed. Sorbent tubes packed with Tenax TA were used to trap styrene vapour purged from urine. Thermal desorption-gas chromatography was used for sorbent tubes analysis. The detection limit (0.70 μg/l), linearity range, recovery (>94% for spike urine samples) and repeatability for urine from occupational exposed workers show the suitability of the method for the determination of styrene in urine. One specific advantage of this method is the possibility of storage of the charged sorbent tubes during long periods of time without a significant loss of styrene. This approach can be used, with slight modifications, for urinary determination of several others organic contaminants commonly present in occupational exposures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.