Abstract

PurposeTo show that for tissues the conspicuous asymmetries in the frequency response function of bSSFP can be mitigated by using a short enough TR.Theory and MethodsConfiguration theory indicates that bSSFP becomes apparently “pure” (i.e., exhibiting a symmetric profile) in the limit of TR →0. To this end, the frequency profile of bSSFP was measured as a function of the TR using a manganese‐doped aqueous probe, as well as brain tissue that was shown to exhibit a pronounced asymmetry due to its microstructure. The frequency response function was sampled using N=72 (phantom) and N=36 (in vivo) equally distributed linear RF phase increments in the interval [0,2π). Imaging was performed with 2.0 mm isotropic resolution over a TR range of 1.5–8 ms at 3 and 1.5 T.ResultsAs expected, pure substances showed a symmetric TR‐independent frequency profile, whereas brain tissue revealed a pronounced asymmetry. The observed asymmetry for the tissue, however, decreases with decreasing TR and gives strong evidence that the frequency response function of bSSFP becomes symmetric in the limit of TR →0, in agreement with theory. The limit of apparently pure bSSFP imaging can thus be achieved for a TR ∼ 1.5 ms at 1.5 T, whereas at 3 T, tissues still show some residual asymmetry.ConclusionIn the limit of short enough TR, tissues become apparently pure for bSSFP. This limit can be reached for brain tissue at 1.5 T with TR ∼ 1–2 ms at clinically relevant resolutions.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.