Abstract

One challenge on the path to delivering FLASH-compatible beams with a synchrotron is facilitating an accurate dose control for the required ultra-high dose rates. We propose the use of pulsed RFKO extraction instead of continuous beam delivery as a way to control the dose delivered per Voxel. In a first feasibility test, dose rates in pulses of up to 600 Gy s−1 were observed, while the granularity at which the dose was delivered is expected to be well below 0.5 Gy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.