Abstract
Studies need to define the optimal parameters under which the holmium laser should operate for arthroscopic meniscectomy. This study was designed to analyze the effect of various Holmium wavelength pulsewidths on human meniscal tissue penetration rates and lateral thermal injury. Using a pulsed Holmium: YAG laser at a wavelength of 2.1 microns, the effect of various pulse-widths on tissue penetration rates as well as the degree of accompanying thermal damage in human meniscal tissue was evaluated in a specially designed jig. Holding the energy constant at 500 mJ per pulse, the pulsewidth was varied between 100 and 600 microseconds. Fiber penetration of meniscal tissue was found to be fastest at a pulsewidth of 250 microseconds. As the pulsewidth was increased or decreased around this number, the observed penetration time decreased, although no statistical difference was found. The size of the hole created was inversely related to the penetration time. Microscopic examination revealed zones of lateral thermal effect extending 800 microns from the ablation site. No relationship between the pulsewidth and the lateral thermal effect could be found.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.