Abstract

PurposePulsed-field ablation (PFA) is a new energy source to achieve pulmonary vein isolation (PVI) by targeted electroporation of cardiomyocytes. Experimental and controlled clinical trial data suggest good efficacy of PFA-based PVI. We aimed to assess efficacy, safety and follow-up of PFA-based PVI in an early adopter routine care setting.MethodsConsecutive patients with symptomatic paroxysmal or persistent atrial fibrillation (AF) underwent PVI using the Farawave® PFA ablation catheter in conjunction with three-dimensional mapping at two German high-volume ablation centers. PVI was achieved by applying 8 PFA applications in each PV.ResultsA total of 138 patients undergoing a first PVI (67 ± 12 years, 66% male, 62% persistent AF) were treated. PVI was achieved in all patients by deploying 4563 applications in 546 PVs (8.4 ± 1.0/PV). Disappearance of PV signals after the first application was demonstrated in 544/546 PVs (99.6%). More than eight PFA applications were performed in 29/546 PVs (6%) following adapted catheter positioning or due to reconnection as assessed during remapping. Mean procedure time was 78 ± 22 min including pre- and post PVI high-density voltage mapping. PFA catheter LA dwell-time was 23 ± 9 min. Total fluoroscopy time and dose area product were 16 ± 7 min and 505 [275;747] cGy*cm2. One pericardial tamponade (0.7%), one transient ST-elevation (0.7%) and three groin complications (2.2%) occurred. 1-year follow-up showed freedom of arrhythmia in 90% in patients with paroxysmal AF (n = 47) and 60% in patients with persistent AF (n = 82, p = 0.015).ConclusionsPFA-based PVI is acutely highly effective and associated with a beneficial safety and low recurrence rate.Graphical abstract

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call