Abstract

We demonstrate an improved technique for nanomechanical imaging in atomic force microscopy. By merging the sensitivity to contact stiffness inherent to contact resonance (CR) spectroscopy with the delicate nature and potential for adhesion data of pulsed-force mode, we address major shortcomings of both techniques. Fast CR data are recorded during each pulsed cycle by driving the sample at two frequencies near the CR frequency and modeling the contact as a harmonic oscillator. The technique provides nanomechanical parameters including frequency, quality factor, and adhesion force. Compared to continuous contact, the technique should reduce damage and support more complex analysis models.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.