Abstract

The authors introduce a new type of silver electrodes. The new electrodes are designed to be disposable and as such they are used only for a limited period of time, usually 5–7 days, and then discarded. Manual polishing, which is required with non-disposable electrodes, is not necessary with disposable electrodes. This manuscript describes the development of pulsed amperometric detection waveforms that make it possible to utilize disposable silver electrodes in the chromatographic detection of sulfide, cyanide, iodide, thiosulfate, bromide and thiocyanate. Also discussed is an evaluation of analytical performance of the electrodes with the newly developed waveforms. The detection limits were found to be 1 ppb (S 2−), 2 ppb (CN −), 5 ppb (Br −, S 2O 3 2− and I −) and 10 ppb (SCN −) for a 10 μl injection volume. Calibration plots for S 2−, CN −, Br − and S 2O 3 2− are linear from their respective detection limits up to 50–100 ppm. The linear calibration for I − and SCN − can be obtained from the detection limits up to 10 ppm. Reproducible long-term performance was confirmed in a 2-week experiment with 10 μl injections of 1 ppm CN −.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.