Abstract

Measurement of systemic blood flow is of crucial importance in patients on mechanical circulatory support (MCS). We reported the case of a 65-year-old female patient in severe cardiogenic shock undergoing left (Jarvik 2000 axial flow pump) and right (Levitronix-Centrimag centrifugal pump) ventricular assist device implant. Evaluation of blood flow was obtained by ultrasonic flowmetry, continuous thermodilution technique, and pressure recording analytical method (PRAM). This pulse contour system allows beat-by-beat systemic blood flow assessment from the analysis of radial artery pressure waveform. At a Jarvik pump speed < or = 10000 rotations per minutes (rpm), thermodilution and PRAM showed similar blood flow values. At a Jarvik pump speed > or = 11000 rpm, the aortic valve did not open and PRAM did not provide blood flow values due to nonpulsatile blood flow. The present paper describes the first experience with PRAM in a single patient on MCS. Further studies are required to assess the validity of PRAM as an additional monitoring system in the setting of ventricular assist device support.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.