Abstract

In this research Seismocardiogram (SCG) signal was used for timing the proximal pulse while pulse-plethysmogram (PPG) was used for timing the dorsal pulse for computing pulse transit time (PTT). Lower body negative pressure (LBNP) was applied for gradually reducing the pulse pressure in 9 subjects. The correlation between the pulse pressure and PTT derived using SCG was studied. It was determined that a significant correlation existed between PTT and pulse pressure for 7 out of the 9 subjects studied with an average significant correlation of 0.88±0.05. The results from this research show a potential application for a system including SCG and PPG to be used in an estimation or trending of pulse pressure.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.