Abstract

Summary form only given. For the new large germanium detector arrays (e.g. EUROGAM, Gammasphere, Euroball, etc.) which are currently under design/in construction, an improved time resolution is desirable to improve the coincidence time resolution. Currently with Ge-Ge timing one can reach a time resolution of the order of 20 ns with large detectors (>30% efficiency). By digitizing the signal every 10 ns with a flash analog-to-digital converter (ADC) system and a following analysis of the digitized pulse, the time resolution can improve relative to conventional techniques (time filter and constant fraction). Measurements show a time resolution of about 9 ns using a conventional technique for a fast plastic and a high-purity germanium (HPGe) detector. With digital pulse shape analysis a resolution of 3 ns is observed. >

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.