Abstract

Pulse oximetry represents a major advance in patient monitoring, but measurement below 70 to 80% saturation has important limitations. Several authors have tested pulse oximetry at low saturations with conflicting results. A review of these data indicates that every patient with a pulse oximeter value below 75 to 80% SaO2 should have one or more invasive measurements of the arterial SaO2 in order to avoid undetected severe hypoxemia.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.