Abstract

The Masimo Radical 7(®) is a new pulse CO oximeter designed to measure methemoglobin. The device has not been evaluated in a clinical setting. In this prospective observational study we compared the arterial methemoglobin levels and the corresponding pulse CO-oximetric values of the Radical 7(®) in regional anesthesia with prilocaine. We analyzed 360 data pairs with methemoglobin values up to 6.6%. The mean bias and limits (± 1.96 sd) of the device were 0.27% (± 1.33%). We found a high degree of agreement in measurement of methemoglobin between the 2 methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.