Abstract

Pulse Oximeter devices are widely used as a non-invasive method for instant monitoring of blood oxygen saturation and heart rate. In this paper, a wireless microcontroller based pulse oximeter is proposed to measure the oxygen delivered to the patient via the oxygen flowmeter. In the first step, the signals received from reusable SpO 2 sensor (finger probe) are processed by a microcontroller to determine the blood oxygen saturation and heart rate. Depending on the current blood oxygen saturation value, wireless signals are sent to the non-invasive ventilation flow meter vacuum regulator to deliver the necessary oxygen into the patient. Oxygen supplied to the patient is automatically controlled according to the oxygen saturation change.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.