Abstract

The main purpose of this study was to develop an implantable direct-electromagnetic left ventricular assist system driven by a linear actuator (linear LVAS). The linear LVAS is a pulsatile pump with a pusher plate that is driven directly by a linear oscillatory actuator (LOA) without any movement converters. This prototype pump unit with a LOA was 100 mm in diameter, 50 mm in thickness, and weighed 740 g. The full-fill/full-eject driving method was applied to the control algorithm. In addition, a mechanism to detect and release sucking was realized to overcome this problem that accompanies the active-filling type of VAS. The performance of the linear LVAS was evaluated in a long-term animal experiment using a goat (56 kg). The goat survived for 42 days. The reason why we terminated this experiment was that thrombus was found in the pump. There was no frictional debris found around the LOA. The linear LVAS did not exhibit electrical or mechanical problems during the first animal experiment.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call