Abstract
BackgroundIn trauma patients, particularly with head immobilization, tracheal intubation without the use of a stylet may be impossible. ObjectivesTo report a very rare but potentially fatal complication that may happen in any Emergency Department: fracture of the plastic sheath of an intubation stylet, reported only twice before in the literature. Case ReportTwo large plastic fragments detached from a stylet while intubating a trauma patient. One piece was removed from the endotracheal tube a few hours later in the operating room. The second fragment migrated asymptomatically into the pulmonary airway. It was successfully retrieved from the right bronchus 24 h later. ConclusionThis potentially life-threatening event may go unnoticed after intubation if the endotracheal tube is not obstructed by the fragment. Gentle withdrawal of the stylet from the tube is essential to avoid stylet fracture. Careful examination of the stylet after intubation may suggest a stylet fracture.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.