Abstract

The occurrence of the pulmonary capillary hemorrhage (PCH) bioeffect of diagnostic ultrasound in rats was investigated for a SuperSonic Imagine shear wave elastography system (Aixplorer, Supersonic Imagine, Aix-en-Provence, France). The elastography imaging repeated at 1 Hz and consisted of widely spaced B-mode image pulses, supersonic push (SSP) pulses and shear wave imaging (SWI) pulses. Groups of rats anesthetized with ketamine and xylazine, or with ketamine only, were imaged on their right side in a warm water bath for one frame, 30 s and 300 s. The image focus and region of interest were adjusted to give exposure only with the background B-mode imaging, or primarily with the SSP and SWI pulses. A sham group had only low power aiming scans. The lungs were removed 5 min after exposure and evaluated for PCH area and volume. The B mode was notably ineffective and produced significant PCH only at the maximum 0 dB output. The SSP pulses together with the SWI pulses produced significant PCH for 300 s, 30 s and even single image exposures. Peak rarefactional pressure amplitude PCH thresholds for 300 s exposure were the same with or without the B-mode pulses at 1.5 MPa (in situ mechanical index, MIIS = 0.67). A 30 s duration resulted in a slightly increased threshold of 1.7 MPa (MIIS = 0.76). The omission of xylazine from the anesthetic, which reduces the sensitivity of rat lung to PCH occurrence, resulted in an increased threshold of 2.1 MPa (MIIS = 0.94). The unique SSP pulses were much more effective than the B mode, but thresholds were comparable to previous results with other diagnostic ultrasound modes on other systems.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call