Abstract
In this article, we present a new method called point spread function (PSF)-Radon transform algorithm. This algorithm consists on recovering the instrument PSF from the Radon transform (in the line direction axis) of the line spread function (i.e., the image of a line). We present the method and tested with synthetic images, and real images from macro lens camera and microscopy. A stand-alone program along with a tutorial is available, for any interested user, in Martinez (PSF-Radon transform algorithm, standalone program). RESEARCH HIGHLIGHTS: Determining the instrument PSF is a key issue. Precise PSF determinations are mandatory if image improvement is performed numerically by deconvolution. Much less exposure time to achieve the same performance than a measurement of the PSF from a very small bead. Does not require having to adjust the PSF by an analytical function to overcome the noise uncertainties.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.