Abstract
We present an in-depth failure analysis of two large diameter bearing metal-on-polymer (MoP) modular total hip replacements, which have required revision surgery due to pseudotumour formation. The failure analysis showed a discrete pattern of material loss from the distal end of the head taper/stem trunnion interface. We postulate that the use of a proximal contacting taper design had provided insufficient mechanical locking between the head and the stem, enabling the head to toggle on the trunnion. In addition, the difference in angle between the taper and the trunnion formed a crevice between the two components. Through a combination of crevice environment, mechanically assisted corrosion, mechanical wear and erosion; debris and metal-ions have been released resulting in the adverse local tissue reactions (ALTR).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.