Abstract

This study evaluates applicability of microbial sensors based on Pseudomonas putida P67.2 (Ps. p.) and Pseudomonas fluorescens P75 (Ps. fl) for BOD measurements in the phenol-containing wastewaters that mimic the wastewaters from oil shale industry. Sensors are calibrated with OECD synthetic wastewater. Linear range in calibration solution for Ps. p. and Ps. fl. sensors is BOD 0-50 mg/L and 0-65 mg/L, respectively. The steady state response time for both sensors is 15-45 min. Measurements show that although Ps. fl. has better sensitivity to calibration solution, the concurrence between sensor-BOD and BOD7 in phenol-spiked wastewater is better for Ps. p. sensor. In linear range the error is only 1.2–13%. For Ps. fl. the error in linear range is 17–55%. Preconditioning in phenol solution (BOD = 5 mg/L) increases sensor’s sensitivity to phenol considerably (52%). This study demonstrates that preconditioned microbial sensors with specifically selected cultures can give better results for fast BOD determination in specific wastewaters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.