Abstract
To the Editor. —Malfunction of a permanent cardiac pacemaker is most often manifested by intermittent or absolute cessation of function or by a reduction from the chronic stable pacing rate. On rare occasions, however, pacemaker tachycardia occurs with rate acceleration up to 100 beats per minute, ie, runaway pacemaker. In this report, pseudo-runaway pacemaker, a previously unrecognized phenomenon, will be presented. Report of Cases.—Case 1. —A 46-year-old woman had a R-wave inhibited demand pacemaker inserted for episodic Stokes-Adams attacks with verified high-degree atrioventricular block. The pacing rate was 72 beats per minute (Figure, A). A routine ECG, taken three months later during an emergency room visit for noncardiac chest pain, showed 100% pacemaker capture at 90 beats per minute (Figure, B). Although the pulse generator was implanted only three months earlier, component failure was suspected, and the unit was replaced. Case 2. —A 70-year-old man had a demand
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.