Abstract

In this work spontaneous oxidation of Prussian white (PW), which is completely reduced form of Prussian blue (PB), has been investigated at ambient conditions and stable form of PW-PB as an intermediate oxidation state based on the presence of both redox forms (PW and PB) was identified and investigated. Two types of optical slides were manufactured by modifying optically transparent indium tin oxide (ITO) coated glass electrode (glass/ITO) (i) with PW layer (glass/ITO/PW optical slide) and (ii) with PW-PB layer (glass/ITO/PW-PB optical slide). Further modification of the developed optical slides with glucose oxidase (GOx) showed that glass/ITO/PW-PB/GOx optical slide was insensitive toward glucose, but glass/ITO/PW/GOx optical slide acted as glucose biosensor and showed linear sensitivity toward glucose in the concentration range from 20 μM to 200 μM (R2 = 0.99).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.