Abstract

Residual fragments remain in over 50% of treatments for lower pole kidney stones. A second‐generation device based on a diagnostic ultrasound system and scanhead has been developed with a unique algorithm for stone detection and the capability to focus ultrasound to expel residual fragments. Focused ultrasound was applied to a bead on string in a water tank as well as to human stones (<5 mm) implanted in the lower pole of a live porcine model via retrograde ureteroscopy. Histological samples were collected and scored in a blinded fashion for therapeutic exposures and for super‐therapeutic levels. The in‐vitro bead was visually observed to move under focused ultrasound. Even with progressive manual displacement of the bead, the system continuously tracked and caused bead movement in real time. In the live porcine model, stones were expelled from the lower pole to the ureteropelvic junction in seconds to minutes using pulses at a duty factor of 0.02 and 8 W total acoustic power. Injury was observed no more frequently than in controls. Occurrence of injury rose slightly above control at a duty factor of 0.02 and 80 W and at a duty factor of 1 and 8 W. [Work supported by NIH DK48331, NIH DK086371, and NSBRI through NASA NCC 9‐58.]