Abstract
A self-tracking ENDOR-induced EPR spectrometer is described, which allows analog correction of the frequency change of the observed ENDOR transition during the magnetic field sweep. Deviations from a linear magnetic field dependence are discussed. New applications in powder samples are given which allow determination of the hfs principal values and axes directions with respect to a non-coaxial g tensor.