Abstract

Two different types of counter-current chromatographs, the cross-axis coil planet centrifuge ( X-axis CPC) and horizontal flow-through coil planet centrifuge (horizontal CPC), were evaluated for protein separation with an aqueous-aqueous two-phase polymer system. The sample solution, containing 10–200 mg each of cytochrome c, myoglobin, ovalbumin and hemoglobin in 2 ml of each phase was eluted with the lower phase. In both instruments, the effects of the flow-rate, revolution speed, and parameter β (helical diameters of the multilayer coil) on the protein separation were investigated. The best results were obtained from the X-axis CPC operated at 750 rpm and a flow-rate of 2.0 ml/min using a multilayer coil with a small helical diameter (β = 0.25–0.60). Four protein samples were well resolved in less than 5 h.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.