Abstract
The lead equivalence of structural barriers surrounding 203 diagnostic x-ray installations was measured and adequacy determined by comparison with the lead thickness specified by the ICRP for distance, nature of occupancy and workload applicable to each installation, assuming operation at 125 kVp. Barrier deficiencies were found in 9%. Protective clothing was also examined: 16 of 314 gloves and 4 of 337 aprons were defective.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.