Abstract

Background A colorectal anastomotic leak can be life-threatening. We have assessed the feasibility of a new intraluminal biodegradable bypass device that we designed to avoid anastomotic leakage and the necessity of a temporary stoma. Methods Fifteen patients underwent colorectal surgery. Before performing the anastomosis in a standard way, the C-Seal (Polyganics BV, Groningen, The Netherlands) was glued to the anvil of the circular stapler. Consequently, the bypass was fixated in the staple row just proximal of the simultaneously made anastomosis. A water-soluble contrast enema was performed after 1 week. Results The sheath was well compatible with the standard stapler device and easy to use. All donuts remained intact. No radiologic or clinical leaks were observed after surgery. Conclusions This pilot study shows that the C-Seal can be applied successfully in colorectal surgery. Further investigation with large numbers of patients is now necessary to assess the efficacy of the C-Seal as a protective device.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.